4.4 ADC Adapter |
An ADC Adapter can be connected to the SP-ICE-3 Card in order to record analogue signals specific to an application.
For process monitoring, in particular, the signal samples from the adapter can be correlated with scanner position data in the time domain.
Such signals could be, for instance:
Workpiece surface temperature, measured by a pyrometer;
Real laser output power level, detected by a suitable sensor;
etc.
The ADC Adapter provides four independent analogue input channels.
Resolution | 16 bits |
Sampling Rate | 100000 per second |
Input Connection | Differential or Single-Ended |
Reference | AGND (Isolated from the SP-ICE-3 Card's digital GND, and from the slot bracket.) |
Output | Signed integer value in the range: -32768 (Uin_diff = -5V) to +32767 (Uin_diff = +5V) . |
The ADC Adapter does not auto-configure itself at power-on (unlike, for instance, the XY2-100 Adapter).
Please refer to 4.4.3 ADC Adapter Configuration.
If the red "Conf Pending" LED is on (see diagram), the adapter has not yet been configured.